Testing Setup for Relative Resistance to Vented Water – Tree
Growth in Solid Dielectric Insulating Materials (As per ASTM D6097-01a)
The System comprises of :
A A) Power
Source for Testing of Solid Insulating :
The equipment is designed to carry out standard test method for Relative Resistance to vented Water–Tree Grown
in Solid Dielectric Insulating Materials as per ASTM D6097-01a.
Ø Specification
:
·
Input Voltage : 3 Phase 415V AC 50Hz 12 Amp.
·
Output Voltage : 500V to 5000V AC (Adjustable
manually) @ 600mA.
·
Output Frequency : 1000 Hz
·
Duty Cycle : Continues
·
Distortion : Less than 5%
·
Output Adjustment : Manually through Poten
tiometer.
Ø
Meter Indication :
·
Voltage : 0 to 10kV AC
·
Current : 0 to 1 Amp
·
Frequency : 50 Hz to 1000 Hz
Ø
Protection :
·
Over current / short circuit with electronic
circuity.
·
ON/OFF control through Zero Interlock.
·
Door Interlock for Test Chamber.
· Emergency Push Button.
B B) Siemens make LOGO PLC for sequence control & testing control.
C C) Water Bath with S.S. material Digital
Temperature to controller to monitoring Temperature between Ambient to 70°C.
D D) Specimen Holder