Single Station High Voltage Inclined Plane Tracking and Erosion Tester - PC Based
Features:
1. PC
Operated through NI DAQ OR SCR make DAQ system with LabVIEW
software.
2. Specially
designed electrodes as per product under test
3. Lighting
arrangement in test chamber
4. Facility
to easily calibrate the system for voltage as well as time calibration.
5. Seven
segment LED display for voltage and current indication .
PC Datalogging comprises of :
1. PC
based Software: This is useful for auto selection of Type of Test, Ranges and
sequence from PC (through RS 232 C Port or Ethernet). This Software (developed
in NI LabVIEW environment) provides the data acquisition on PC. Software
features include user hierarchy and password protection, preconfigurable test
recipes and setting, automatic and quick selection of a test recipe with
auto-upload of values, actual test and result screens with waveforms, test
reports exportable to excel formats, diagnostic checking of the test bench
system, etc.
2. PC Either Dell or HP make with OS will be Windows 10 license.
Specifications:
1. Input supply : 1 phase 230 V AC, 50hz
2. Output voltage : Regulated 100 V to 7500V, @100mA AC programmable.
3. Series ballast selection – 1k, 10k,20k and 50k with link.
4. Tripping current setting : 60 mA
5. Electrode system : Stainless still material.
6. Angle of inclination : 45 deg.
7. Fixture arrangement: Made for flat test specimen as well as for various sizes of cable.
8. Liquid deliver capacity: 0.075 ml/ min to 0.900 ml/ min, Programmable
9. Test chamber : UV protected Glass door with door interlock.
10.Enclosure : Built with aluminum profile and MS covers with powder coating
11. Test Chamber : 50cm X 25cm X 25cm(L X H X D)