GWT A3 ( PC based system)- In this variant, setting will be done through PC software and report can be generated.
Salient Features:
1. A motorized sliding carrier with an universally adjustable SS perforated tray is provided to mount the specimen.
2. Self adjusting current source to heat & maintain the glow wire to the preset temperature is provided.
3. The mounting jig can be moved forward or reverse with the help of a push button enabling ease of operation throughout the test.
4. Thyristor based proportional controller with ambient temperature compensation (100 deg. C min. ramp) is provided to get excellent control.
5. Precision scale is provided to measure the height of the flame.
6. Lamp Indications for Mains ON & Test ON.
7. Safety – Emergency Switch, MCB, Fuses.
8. External Calibration terminals for Temperature and Time.
9. Test Lock key switch to ON/OFF or avoid unauthorized person to operate the equipment.
10.Push buttons are provided to record the Ignition Time and Extinguish Time.
Specifications:
1. Input Supply: – 230 V AC, 50 Hz, 1 Phase, 500 VA.
2. Micro-controller based programmable controller (MTCM Module) – The setting of temperature (Ambient to 960 deg C) and time (0 to 99.9 sec) is through keypad.
3. Test Current: Between 120A to 150A
4. Temperature Sensor – Thermocouple: Fine wire of Cr/Al of 1 mm diameter
5. Seven Segment LED single display – For temperature and current indication.
6. Heating Element – Glow Wire: Nickel / Chromium (Ni/Cr-80:20) wire of 4 mm diameter with a shape as per the standard.
7. Penetration Depth: The limit of penetration of the glow wire is automatically restricted to 7 mm.
8. Test specimen contact force against the glow wire is spring-loaded to 0.8 to 1.2N
Variants & Customization:
1. Lab view test ware to control sequence of operation.
2. Setting and control of Heating, Control of current will be through PC.
3. Test Time, Application time will be settable. PC will record Ignition Time & Extinguish Time with manual input from operator.
4. PC Either Dell OR HP make. OS will be Windows 8 license.
5. NI interface system.
6. Speed of carriage of specimen will be adjustable as per standard through PC.
7. Results will be stored on PC as per test samples & date.